CF11: The Laboratory Reference for Continuous Monitoring

The Laboratory Reference for Continuous Monitoring The CF11 by Air Leak Control is a high-precision measuring instrument exclusively designed for laboratory applications and R&D. It was developed to provide a complete and reliable testing platform, ensuring optimal metrological-quality results.

  • Description
  • Features

The CF11 integrates, within a single compact enclosure, all the critical instrumentation required for complete characterization: Pressure sensor Temperature sensor Dual flow sensor (with selectable measurement ranges to suit your specific needs) The CF11 operates in continuous test mode, enabling uninterrupted monitoring and real-time data analysis. To ensure maximum accuracy and full traceability of results, the device incorporates an advanced thermal protection system. Built-in protection against excessively high or low ambient temperatures ensures that measurements are always carried out under optimal conditions. The CF11 features a simplified and ergonomic user interface. Its intuitive design makes operation easy, even for complex test protocols, allowing laboratory personnel to focus on result analysis rather than instrument setup.

Pressure RangeResolutionUncertainty (Leak)
0-500 mbar0,1 mbar+/-1% + 1 digit
0-5 bar1 mbarIdentical
0-10 bar1 mbarIdentical
Flow RangeResolutionUncertainty (Leak)
1,20 l/h0,01 l/h +/- 2,5% +1 digit
12 l/h0,1 l/hIdentical
150 l/h0,1 l/hIdentical
2000 l/h1 l/hIdentical
5000 l/h1 l/hIdentical
DisplayLCD screen, 4 lines × 20 characters
Operating Temperature+5 to +60 °C
Storage Temperature-15 to +75°C
HumidityMaximum 20% (non-condensing)
Power Supply Required+24VDC (1.5 A minimum)
Power SupplyAC power adapter included
Dimensions (H × W × D)135 x 170 x 120 mm
WeightApprox. 2kg

Maintenance and Accessories

  • Entretien Recommandé :
    • Étalonnage annuel obligatoire pour la certification de précision.
  • Adaptable Accessories (optional):
    • Remote display